Walk-In-Stability Test Chamber


    Control SystemProgrammable Logic Control with Safety controller, auto changeover of Standby systems & other event managements with 3.8” Color HMI with touch screen and Ethernet connectivity
    Air CirculationBlowers to have uniformity of Temperature under loaded condition.
    AlarmAudio Visual Alarm if the temperature deviates from the preset Temperature
    LightingInterior illumination for working area.
    Insulation80mm Thick CFC free PUF Insulated modular cabinet
    Heating SystemU shaped S.S. Nichrome wire air heaters
    Observation DoorInside see through unbreakable acrylic door
    Refrigeration SystemHermetically Sealed CFC Free Emerson Compressor with R- 22 refrigerant having state of art By pass system for continues run compressor
    Humidity SystemNon condensing type steam injection system with water level arrangement with insulation to save heat energy.
    PowerWorks on 230/430 V AC single / three phase 50Hz
    OtherMCB, Heavy duty latch with lock & key.
    Safety Features
  1. Float switch to cutoff the mains supply in case of low water level in the boiler tank.
  2. Additional safety thermostat to cut off the boiler system in case of overshoot of temperature.
  3. 180 Sec. compressor ‘ON’ delay timer to safe guard the compressor.
    Certifications ISO 9001: 2008 and ISO 13485 : 2012